Teledyne API T703
The J.J. Wilbur Company is proud to offer the Teledyne API T703. A microprocessor controlled ozone calibrator for precision gas analyzers. The Model T703 can deliver concentrations from 50 ppb to 5 ppm at flows from 1 LPM to 5 LPM.
The Model T703 has two major components; the ozone generating device and the photometer. The ozone generator produces ozone by passing zero air over a UV lamp. Flow rate and lamp intensity are adjusted to deliver a specified ozone concentration by volume. The temperature and pressure compensated photometer, identical to that of the Model T400 ozone analyzer, provides feedback control of the UV lamp intensity, assuring stable ozone output. A nearly unlimited number of calibration sequences may be programmed into the non-volatile memory of the T703, covering time periods up to a year.
All T Series instruments offer an advanced color display, capacitive touch screen, intuitive user interface, flexible I/O, and built-in data acquisition capability. All instrument set up, control and access to stored data and diagnostic information is available through the front panel, or via RS232, Ethernet, or USB com ports, either locally or by remote connection.
The Model T703 comes with NumaView™ Software. NumaView™ Remote PC Software allows for a remote connection with virtual interface and data downloading capability to analyzers operating NumaView™ Software.
Product Benefits
Primary or transfer standard for calibration of ozone analyzers
Large, vivid, and durable color graphics display with touch screen interface
Ethernet, RS-232, and (optional) USB com ports
Front panel USB connections for peripheral devices and firmware upgrades
12 independent timers for sequences
Nested sequences (up to 5 levels)
UV Lamp Feedback modes: current control; photometer control
Internal zero air pump (optional)
Flow adjustable from 1 LPM to 5 LPM
High Flow/High Output capability (optional)
Inlets for external ozone reference sources
Two-year warranty